Silicon photomultipliers (SiPMs) were considered as photodetectors in Cherenkov time-of-flight positron emission tomography (TOF PET) due to their relatively high photon detection efficiency (PDE) which resulted in a higher gamma detection efficiency of the Cherenkov PET detector. SiPMs produced by KETEK, AdvanSiD, SensL and Hamamatsu were tested in this application. The coincidence time resolution, gamma detection efficiency and single side event rates were measured in a simple experimental setup, consisting of two detectors placed in a coincidence configuration. The apparatus was cooled to -25° C in order to reduce the SiPM dark count rates. A single side gamma detection efficiency of up to 26% was measured, which is a marked improvement over previous Cherenkov TOF PET measurements. The best measured coincidence time resolution of 297 ps FWHM was limited by the time response of the SiPMs used for this experiments.
The performance of silicon photomultipliers in Cherenkov TOF PET
R. Dolenec,S. Korpar,P. Križan,Rok Pestotink
Published 2015 in IEEE Transactions on Nuclear Science
ABSTRACT
PUBLICATION RECORD
- Publication year
2015
- Venue
IEEE Transactions on Nuclear Science
- Publication date
2015-05-18
- Fields of study
Medicine, Physics
- Identifiers
- External record
- Source metadata
Semantic Scholar, PubMed
CITATION MAP
EXTRACTION MAP
CLAIMS
- No claims are published for this paper.
CONCEPTS
- No concepts are published for this paper.
REFERENCES
Showing 1-3 of 3 references · Page 1 of 1
CITED BY
Showing 1-20 of 20 citing papers · Page 1 of 1