Fluorescent probes are a popular and indispensable tool for monitoring sodium concentration in living cells in situ. Calibration of fluorescent probes inside cells commonly uses ionophores to equilibrate intracellular and external ion concentrations. Here we test this calibration method using in parallel classical flame emission assay. Suspension human lymphoma cells allow both flow cytometry fluorometric study and flame emission assay. The most sensitive Na+ fluorescent probe ANG-2 and the most common ionophores were tested. Cellular Na+ was altered for calibration in three different ways: by stopping the sodium pump with ouabain, by inducing of apoptosis with staurosporine, and by gramicidin or amphotericin B treatment. We found that ANG-2 fluorescence in cells treated with gramicidin or amphotericin was about two fold lower than in the cells with the same sodium concentration but without ionophores. The equal fluorescence measured in the absence and in the presence of ionophores corresponds to different cell sodium concentrations. No effect of gramicidin on hydrolyzed ANG was observed in vitro. The mechanism, by which gramicidin decreases ANG fluorescence in cells is unlikely to be physical quenching and remains obscure. We conclude that ANG fluorescence does not display realistic cell Na+ if fluorescence in cell is measured in ionophore absence while calibrated in its presence.

PUBLICATION RECORD

  • Publication year

    2019

  • Venue

    bioRxiv

  • Publication date

    2019-07-26

  • Fields of study

    Biology, Medicine, Chemistry

  • Identifiers
  • External record

    Open on Semantic Scholar

  • Source metadata

    Semantic Scholar

CITATION MAP

EXTRACTION MAP

CLAIMS

  • No claims are published for this paper.

CONCEPTS

  • No concepts are published for this paper.

REFERENCES

Showing 1-32 of 32 references · Page 1 of 1

CITED BY

  • No citing papers are available for this paper.

Showing 0-0 of 0 citing papers · Page 1 of 1