Implantable functional electrical stimulation (IFES) has demonstrated its effectiveness as an alternative treatment option for diseases incurable pharmaceutically (e.g., retinal prosthesis, cochlear implant, spinal cord implant for pain relief). However, the development of IFES for gastrointestinal (GI) tract modulation is still limited due to the poorly understood GI neural network (gut–brain axis) and the fundamental difference among activating/monitoring smooth muscles, skeletal muscles and neurons. This inevitably imposes different design specifications for GI implants. This paper thus addresses the design requirements for an implant to treat GI dysmotility and presents a miniaturized wireless implant capable of modulating and recording GI motility. This implant incorporates a custom-made system-on-a-chip (SoC) and a heterogeneous system-in-a-package (SiP) for device miniaturization and integration. An in vivo experiment using both rodent and porcine models is further conducted to validate the effectiveness of the implant.
A Wireless Implant for Gastrointestinal Motility Disorders
Y. Lo,Po-Min Wang,Genia Dubrovsky,Ming-Dao Wu,Michael Chan,J. Dunn,Wentai Liu
Published 2018 in Micromachines
ABSTRACT
PUBLICATION RECORD
- Publication year
2018
- Venue
Micromachines
- Publication date
2018-01-01
- Fields of study
Medicine, Engineering
- Identifiers
- External record
- Source metadata
Semantic Scholar, PubMed
CITATION MAP
EXTRACTION MAP
CLAIMS
- No claims are published for this paper.
CONCEPTS
- No concepts are published for this paper.
REFERENCES
Showing 1-38 of 38 references · Page 1 of 1
CITED BY
Showing 1-28 of 28 citing papers · Page 1 of 1