PURPOSE We developed a dynamic B0 shimming approach using a 46-channel AC/DC shim array to correct phase errors caused by eddy currents from diffusion-encoding gradients in diffusion-prepared MRI, enabling high b value imaging without the SNR loss from the use of a magnitude stabilizer. METHODS A 46-channel AC/DC shim array and corresponding amplifier system were built. Spin echo prescans with and without diffusion preparation were then used to rapidly measure eddy current-induced phase differences. These phase maps were used as targets in an optimization framework to compute compensatory shim currents for multi-shot 3D diffusion-prepared acquisitions. RESULTS The proposed method allows flexible use of the AC/DC shim array to correct undesirable eddy current effects in diffusion-prepared MRI. Phantom and in vivo experiments demonstrate whole-brain, cardiac-gated, multi-shot 3D diffusion-prepared imaging without the use of magnitude stabilizers. The approach enables preservation of full SNR while achieving reliable diffusion encoding at b values up to 2000 s/mm2. CONCLUSIONS This work demonstrates a new strategy for applying an AC/DC shim array to compensate for eddy current-induced phase errors in diffusion-prepared MRI. By eliminating the need for a magnitude stabilizer, it enables efficient high-quality diffusion imaging with full signal sensitivity retained.
A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion-Prepared MRI Acquisition Using a Multi-Coil AC/DC Shim-Array.
C. Liao,Jason P. Stockmann,Zhitao Li,Zhixing Wang,Mengze Gao,Lincoln Craven-Brightman,M. Śliwiak,Charles Biggs,Jack Glad,Jiazheng Zhou,Yurui Qian,Zheng Zhong,Nan Wang,Hua Wu,Thomas Grafendorfer,Fraser Robb,Bernhard Gruber,A. Mareyam,Adam B. Kerr,Xiaozhi Cao,K. Setsompop
Published 2025 in Magnetic Resonance in Medicine
ABSTRACT
PUBLICATION RECORD
- Publication year
2025
- Venue
Magnetic Resonance in Medicine
- Publication date
2025-11-09
- Fields of study
Medicine, Physics, Engineering
- Identifiers
- External record
- Source metadata
Semantic Scholar
CITATION MAP
EXTRACTION MAP
CLAIMS
- No claims are published for this paper.
CONCEPTS
- No concepts are published for this paper.
REFERENCES
Showing 1-74 of 74 references · Page 1 of 1
CITED BY
- No citing papers are available for this paper.